    Common faults and causes of zirconia oxygen analyzers

    發布時間: 2024-12-13  點擊次數: 865次

    1. System maintenance.

    If the oxygen indicator shows zero or no oxygen indication in the oxygen quantity converter, first check whether the oxygen indication circuit open. If the circuit is normal, check whether the zirconia probe is normal. At a temperature of 700℃, measure the DC millolt value of the oxygen cell. If the potential is greater than 1 millivolt and the resistance is greater than 100Ω, it indicates the probe is aged and needs to be replaced with a sensor or probe. If the potential is less than 1 millivolt and the resistance is less than100Ω, it indicates an open circuit in the wiring, which should be checked and handled one by one.

    2. constant temperature system failure.

    (1) If the heating temperature cannot be maintained, check whether there is a short circuit or open circuit in the resistance wire inside the heating furnace The normal resistance of the heating furnace wire is about 60-80Ω. If the resistance wire is open or the resistance measured by the multimeter less than 60Ω, the heating furnace wire should be replaced.

    (2) When the converter shows that the temperature is too high or out of control, it not only causes inaccurate oxygen display but also acceler the aging or damage of the zirconia head. At this point, you should immediately check if the thermocouple is open-circuited, or a multimeter to check the thermocouple. If the resistance is greater than 20Ω, it is judged that the thermocouple is open-cuited, and the thermocouple and zirconia head should be replaced; another reason for the temperature失控is: relay or temperature control circuit failure, relay or temperature control circuit failure will cause the oxygen converter to always be in one of the heating voltages of the heating furnace, thus leading to an increase in.

    3. oxygen cell fault.

    (1) After the oxygen cell has been used for a period of time, the contact points of its leads may become oxidized, resulting in contact or detachment. This can cause the Em value (in units of a multimeter) to fluctuate and become unstable. Observe the oxygen cell and take action.

    (2) When the oxygen signal deviation is too large or inaccurate, it may be due to a short circuit in the zirconia electrode, which case the wiring of the zirconia cell should be checked for short circuits. When the temperature does not rise, if the internal resistance is less than 0Ω, it indicates a short circuit in the electrode, and the zirconia head needs to be replaced.

    (3) During the operation of the oxygen measurement device, if the background potential exceeds 10mV or is less than -30mV, it must be recalibrated. If the background potential remains abnormal after calibration, it indicates that the probe has aged and can no longer be used.

    4. pneumatic circuit failure.

    (1) The zirconia head is clogged with ash, causing an obstruction in the reference gas circulation. This results in a decrease in probe output potential Em and an increase in the background potential Eo. In this case, blowing several times with a bulb syringe at the gas inlet of the will resolve the issue after the gas line is unclogged.

    (2) Oxygen cell ash blockage: The probe should be removed from the ash blockage and the filter should be reinstalled.

    (3) Probe leakage has a significant impact on the probe.

    5. The oxygen content on both sides of the boiler is too high.

    By adjusting the boiler combustion conditions, when starting or stopping the boiler fan or coal mill group, if the oxygen content deviation on both sides of the boiler is large, it is a normal phenomenon. Once the boiler combustion conditions stabilize, the oxygen content display will return to normal. If the unit adjustment is not in place and the deviation is large on both sides, it is necessary to check if the zirconia head is working properly by dismantling the transmitter for inspection. The device requires the operator to adjust the gas flow. The gaseous and reference state gases should continuously flow through the electrode without stopping, and there should be no dead ends. However, too large a flow will have a cooling effect, making the cell temperature uneven and causing errors. If the flow is too small, it is not only easy to form stagnant gas flow and dead zones but also may cause local changes in oxygen concentration due to the conduction of oxygen ions, thus affecting the accuracy of the measurement and causing deviations on both sides.

    6. On-site interference causes failures.

    For newly installed instruments, due to the strong electromagnetic field, when an unstable fault occurs, the first thing to consider is the interference issue.


在線客服 聯系方式 二維碼

服務熱線

0550-2409223

掃一掃,關注我們

皖公網安備 34118102000850號

主站蜘蛛池模板: 亚洲国产成人在线视频 | 亚洲国产成人一区二区在线 | 全黄h全肉细节文在线观看 全黄H全肉细节文短篇 | 青柠电影在线看 | 97成人精品视频在线播放 | 9420高清完整版在线电影免费观看 | 穿白丝袜边走边尿白丝袜 | 99久久国产极品蜜臀AV酒店 | 美女被j进去动态 | 成人18视频在线观看 | 老外的好大c的我好爽 | 成人在线小视频 | 国产精品视频成人 | 国产精品日韩欧美一区二区三区 | 男人免费网站 | 全黄H全肉禁乱公 | 我的美女奴隶 | 99RE8国产这里只有精品 | 国产VA精品午夜福利视频 | 精品一区二区免费视频蜜桃网 | 久久精品99热超碰 | 99久久精品费精品蜜臀AV | 国产精品午夜福利在线观看 | av亚洲2017色天堂 | 爆操日本美女 | 中文字幕中文字幕永久免费 | 日本69xx 老师 | 国产在线午夜 | 野花香在线观看免费观看大全动漫 | 国产午夜精品美女免费大片 | 午夜爽喷水无码成人18禁三级 | 10分钟免费观看视频 | 亚洲一区精品伊人久久伊人 | 久久中文字幕综合不卡一二区 | 国产露脸无码A区久久蘑菇 国产露脸无码A区久久 | 撅高 自己扒开 调教 | brazzers情欲狂欢 | 欧式午夜理伦三级在线观看 | 美女被强奷到抽搐的动态图 | yellow片在线观看免费观看动漫 | 综合一区无套内射中文字幕 |