    您當前的位置:首頁 > 技術文章 > Common faults and causes of zirconia oxygen analyzers

    Common faults and causes of zirconia oxygen analyzers

    發布時間: 2024-12-13  點擊次數: 706次

    1. System maintenance.

    If the oxygen indicator shows zero or no oxygen indication in the oxygen quantity converter, first check whether the oxygen indication circuit open. If the circuit is normal, check whether the zirconia probe is normal. At a temperature of 700℃, measure the DC millolt value of the oxygen cell. If the potential is greater than 1 millivolt and the resistance is greater than 100Ω, it indicates the probe is aged and needs to be replaced with a sensor or probe. If the potential is less than 1 millivolt and the resistance is less than100Ω, it indicates an open circuit in the wiring, which should be checked and handled one by one.

    2. constant temperature system failure.

    (1) If the heating temperature cannot be maintained, check whether there is a short circuit or open circuit in the resistance wire inside the heating furnace The normal resistance of the heating furnace wire is about 60-80Ω. If the resistance wire is open or the resistance measured by the multimeter less than 60Ω, the heating furnace wire should be replaced.

    (2) When the converter shows that the temperature is too high or out of control, it not only causes inaccurate oxygen display but also acceler the aging or damage of the zirconia head. At this point, you should immediately check if the thermocouple is open-circuited, or a multimeter to check the thermocouple. If the resistance is greater than 20Ω, it is judged that the thermocouple is open-cuited, and the thermocouple and zirconia head should be replaced; another reason for the temperature失控is: relay or temperature control circuit failure, relay or temperature control circuit failure will cause the oxygen converter to always be in one of the heating voltages of the heating furnace, thus leading to an increase in.

    3. oxygen cell fault.

    (1) After the oxygen cell has been used for a period of time, the contact points of its leads may become oxidized, resulting in contact or detachment. This can cause the Em value (in units of a multimeter) to fluctuate and become unstable. Observe the oxygen cell and take action.

    (2) When the oxygen signal deviation is too large or inaccurate, it may be due to a short circuit in the zirconia electrode, which case the wiring of the zirconia cell should be checked for short circuits. When the temperature does not rise, if the internal resistance is less than 0Ω, it indicates a short circuit in the electrode, and the zirconia head needs to be replaced.

    (3) During the operation of the oxygen measurement device, if the background potential exceeds 10mV or is less than -30mV, it must be recalibrated. If the background potential remains abnormal after calibration, it indicates that the probe has aged and can no longer be used.

    4. pneumatic circuit failure.

    (1) The zirconia head is clogged with ash, causing an obstruction in the reference gas circulation. This results in a decrease in probe output potential Em and an increase in the background potential Eo. In this case, blowing several times with a bulb syringe at the gas inlet of the will resolve the issue after the gas line is unclogged.

    (2) Oxygen cell ash blockage: The probe should be removed from the ash blockage and the filter should be reinstalled.

    (3) Probe leakage has a significant impact on the probe.

    5. The oxygen content on both sides of the boiler is too high.

    By adjusting the boiler combustion conditions, when starting or stopping the boiler fan or coal mill group, if the oxygen content deviation on both sides of the boiler is large, it is a normal phenomenon. Once the boiler combustion conditions stabilize, the oxygen content display will return to normal. If the unit adjustment is not in place and the deviation is large on both sides, it is necessary to check if the zirconia head is working properly by dismantling the transmitter for inspection. The device requires the operator to adjust the gas flow. The gaseous and reference state gases should continuously flow through the electrode without stopping, and there should be no dead ends. However, too large a flow will have a cooling effect, making the cell temperature uneven and causing errors. If the flow is too small, it is not only easy to form stagnant gas flow and dead zones but also may cause local changes in oxygen concentration due to the conduction of oxygen ions, thus affecting the accuracy of the measurement and causing deviations on both sides.

    6. On-site interference causes failures.

    For newly installed instruments, due to the strong electromagnetic field, when an unstable fault occurs, the first thing to consider is the interference issue.


在線客服 聯系方式 二維碼

服務熱線

0550-2409223

掃一掃,關注我們

皖公網安備 34118102000850號

主站蜘蛛池模板: 久久久高清国产999尤物 | 欧洲另类一二三四区 | 久久噜国产精品拍拍拍拍 | 国产AV一区二区三区日韩 | 国产精品久久久久久久A片冻果 | 国产午夜精品一区二区 | 羞羞影院午夜男女爽爽免费 | 国产精品99久久久久久AV蜜臀 | 伊人不卡久久大香线蕉综合影院 | 亚洲国产日韩欧美高清片a 亚洲国产日韩a精品乱码 | 青青草在现线免费观看 | 99草在线观看 | 嫩草欧美曰韩国产大片 | 视频一区国产第一页 | 少妇无码太爽了视频在线播放 | 国产午夜婷婷精品无码A片 国产午夜视频在永久在线观看 | 收集最新中文国产中文字幕 | 毛片在线全部免费观看 | 国产精品乱人无码伦AV在线A | 国产精品久久久久久久久无码 | 娇小老少配xxxxx | 国产一区在线观看免费 | 久久亚洲电影www电影网 | 黄色大片aa | 影音先锋xfplay影院av | 国产成人综合高清在线观看 | 男子扒开美女尿口做羞羞的事 | 夜色福利院在线看 | 91久久综合精品国产丝袜长腿 | 99re久久免费热在线视频手机 | 在线视频一区二区三区在线播放 | 国产麻豆AV伦 | 国产自产视频在线观看香蕉 | 欧洲xxxxx | 禁漫H天堂免费A漫 | 久久伊人电影 | 亚洲国产精品第一影院在线观看 | 久久精品热老司机 | 天天躁人人躁人人躁狂躁 | 男女亲吻摸下面吃奶视频 | 国产亚洲精品久久无亚洲 |